OVERVIEW
FEATURES
MORE INFO
CuattroDR
Arcus - Mobile DR Solution
  • Interface:
  • The CuattroDR™ touchscreen wireless acquisition tablet can be easily removed, allowing it and the wireless detector to be shared with other portable X-ray devices. They can also be deployed as an in-room imaging solution with permanent X-ray installations. This flexibility makes next day loaner replacement possible (if necessary), substantially reducing down time due to system repairs. This is in contrast to integrated on-board CPU workstations that require on-site technician diagnosis and remotely deployed parts for repair.
  • Generator:
  • Constant Potential KW ratings: 20, (32, 40 and 50 available)
  • Kvp Range in 1Kvp Range in 1 KVp steps: 40 to 125 (150 optional)
    Accuracy: +/- (4% + 1mA)
  • Exposure time range: 0.001 - 10 sec.
    Accuracy: +/-(2% +0.1ms)
  • mAs range: 0.1 - 500mAs
  • Operation Independent from mains supply: Standard
  • Collimator: Manual with electronic timer and meter
  • Anatomical programmer: 492 views
  • Automatic exposure control: Optional
  • Connection for Bucky: Optional
  • Remote Wireless: Optional
  • Line voltage range and phases: Single - Phase line automatic regulation from 100 - 240 Vac (+/-10%)
  • Motor Driven Unit: Standard
  • Weight: 541Kg
14x17" Wireless Detector
  • Sensor:
  • Panel: Single substrate amorphous silicon active TFT/diode array
  • Scintillator: Direct deposition CsI:Tl
  • Pixel Matrix: 3556 × 4320
  • Pixel Pitch: 100 μm
  • Electronics:
  • Amplifiers: Low noise ASICs with user selectable gains
  • ADC: 16 bit
  • Image Transfer Time: Wired: 500 ms; Wireless: 3000 ms
  • On Board Memory: 1 GB DDR3, 4 GB SDHC card
  • Mechanical:
  • Size: ISO 4090 for 35 cm × 43 cm (14” × 17”) cassette size
  • Active Area: True 355 mm × 432 mm
  • External Dimensions: 384 mm (w) × 460 mm (l) × 15 mm (h)
  • Weight: 3.8 kg (8.4 lbs)
  • Housing: Carbon-fiber front & back
  • Communications:
  • Status Display: OLED display with WiFi, battery, and sensor indicators
  • Wireless Data I/F: 802.11n WiFi standard
  • Wired Data I/F: GigE via optional power & communication tether
  • X-ray I/F: Integrated X-ray trigger control. Automatic Exposure Detection
  • Imaging Performance:
  • Limiting Resolution: 5 cy/mm
  • Typical MTF: 70% (1 cy/mm), 40% (2 cy/mm), 15% (4 cy/mm) for RQA5
  • Typical DQE: 75% (0 cy/mm), 60% (1 cy/mm), 40% (3 cy/mm) for RQA5
  • Environmental:
  • Temperature: 10 – 35 °C operating
  • Humidity: 30 – 70 % RH operating (non-condensing)
  • Power:
  • Battery: Rechargeable battery, 53.3 Wh
  • Battery Charger: External two bay charger 100-240 V AC 50/60 Hz
  • Interface & Power Unit: Optional XRpad IPU with external power supply - 100-240 V AC, GigE, and X-ray I/F